YL9100 HPLC

  • Ultra speed data processing by Network HPLC
  • YL-Clarity designed for all you need in the lab
  • Enhanced superior performance
  • Versatile dedicated applications
  • Accurate and reliable data
  • Stylish and compact design
Quaternary Pump
  • Operating principle:parallel dual-plunger pump,Low-pressure gradient
  • Number of Solvent:up to 4 solvents
  • Gradient formation:4-channel mixing valve
  • Composition Precision:< 0.1%
  • Composition Accuracy:< 0.5%
Binary Pump
  • Operating principle:Double parallel dual-plunger pump, High-pressure gradient
  • Configure with Gradient pump
Vacuum Degasser
  • Number of channel: 4 channels
  • Maximum flow rate: 10ml/min per channel
  • > 0~2.0ml/min per channel for 70% Gas Removed from Methanol
  • Internal volume per channel: 925 ul per channel
  • Materials in contact with solvent :Teflon AF,PEEK and Glass-filled PTEE
Column Compartment
  • Temperature range: 4°C(Cooling) – 90°C
  • Temperature stability: ± 0.05°C
  • Temperature accuracy: ±0.5°C
  • Column capacity: Max. 4 ea of 30 cm column(Analytical), Max. 2 ea of 30 cm column(Semi-Prep)
UV/Vis Detector
  • Wavelength Range: 190~900nm
  • Data collection rate:up to 50Hz
  • Noise level: < ±0.5 × 10-5 AU ,254nm,dry cell
  • Drift: < 1 × 10-4 AU/hr
  • Linearity: > 99.5% for 2.5AU(acetone,254nm)
PDA Detector
  • No. of PDA Channel:1024
  • Wavelength: 190 ~ 950nm
  • Noise Level: 2 × 10-5 AU(Empty Cell,1 sec Rise Time,254nm)
  • Drift: 2 × 10-4 AU/hr(Baseline Correction), 0.001AU/hr(Room Temp)
  • Wavelenth Accuracy: 1nm (HY-1 Holmium Oxide Filter)
RI Detector
  • RI Range: 1.00~1.75 RIU
  • Noise: ≤ 1.3 × 10-8 RIU
  • Drift: 0.8 × 10-7 RIU/hr
  • Linear dynamic range: 80 × 10-5 RIU